600NDIR

Model 600 NDIR/O2

DESCRIPTION

The California Analytical Models 601-2-3 NDIR/O2 digital analyzer product line is designed around a state of the art 16 bit microprocessor, with 16 digital inputs, 16 digital outputs, 16 analog inputs and 4 analog outputs. The analyzer can be manually operated from the keypad or remotely via TCP/IP, RS-232C communications and discrete inputs. The analyzer display includes screen presentation of all analyzer alarms. Four levels of password protection are provided. For precision measurements, the analyzer’s accuracy is increased by entering calibration curve fit polynomials. Automatic calibration may be activated locally or remotely and includes auto cal via preset times.

METHOD OF OPERATION

The California Analytical Instruments’ NDIR analyzer is based on the infrared absorption characteristics of gases. Using a single infrared beam to measure gas concentrations, this analyzer produces highly stable and reliable results. A single infrared light beam is modulated by a chopper system and passed through a sample cell of predetermined length containing the gas sample to be analyzed. As the beam passes through the cell, the sample gas absorbs some of its energy. The attenuated beam (transmittance) emerges from the cell and is introduced to the front chamber of a two-chamber infrared microflow detector. The detector is filled with the gas component of interest and consequently the beam experiences further energy absorption. This absorption process increases the pressure in both of the chambers. The differential pressure between the front and rear chambers of the detector causes a slight gas flow between the two chambers. This flow is detected by a mass-flow sensor and is converted into an output signal.

METHOD OF OPERATION - OXYGEN

The California Analytical oxygen analyzer section utilizes the paramagnetic method to determine the percent level of oxygen contained in the sample gas. The oxygen level is displayed on the LCD panel in engineering units.

FEATURES


  • Measures from 0-50 ppm up to 100% Full Scale (CO / CO2 / CH4 / C3H8 / SO2)
  • Measures Oxygen (O2) from 0–1% up to 100%
  • Multi-Component
  • Four User Definable Ranges per Component
  • Fast Response Time
  • Auto Ranging / Auto Calibration
  • Comprehensive Diagnostics
  • Temp. & Barometric Pressure Compensation
  • Output Options: Analog (User Scalable), (RS232) using AK Protocol & TCP/IP
  • Displays & Outputs: Concentrations, Diagnostics, Alarms & Preventative Maintenance
  • Remote Monitoring & Control

  • APPLICATIONS

  • Stack Gases (CEM)
  • Combustion Efficiency
  • Turbine/Generator Feedback Control
  • Process Chemical Gas Analysis
  • Personnel Safety
  • Fuel Cell Analysis
  • Vehicle Emission

  • Model 600P Oxygen Only
    Model 601 Any 1 NDIR Component
    Model 602 Any 2 NDIR Components
    Model 603 Any 3 NDIR Components
    Model 601P Any NDIR Component & O2
    Model 602P Any 2 NDIR Components & O2

    SPECIFICATIONS

    IR ANALYSIS METHOD Non-Dispersive Infrared (NDIR)
    NDIR COMPONENTS CO / CO2 / CH4 / C3H8 / SO2
    DETECTOR TYPE Microflow
    NDIR RANGES From 0-50 ppm up to 0-100 % (Dependent upon gas & cell length)
    RANGE RATIO 10:1 (Highest Range/10 = Lowest Range)
    RESPONSE TIME (IR) T90 < 2 Seconds to 60 Seconds Adjustable (Depending on configuration)
    IR SAMPLE CELL Stainless Steel with Replaceable Gold Cell Liner
    RESOLUTION 0.0001 (Displays Five Significant Digits)
    REPEATABILITY Better than 1.0% of Full Scale
    LINEARITY Better than 0.5% of Full Scale of Factory Calibrated Ranges
    NOISE Less than 1% of Full Scale of Factory Calibrated Ranges
    ZERO & SPAN DRIFT Less than 1% of Full Scale per 24 Hours
    ZERO & SPAN ADJUSTMENT Via front panel, TCP/IP or RS-232
    SAMPLE FLOW RATE 0.25 to 2.0 LPM (Consult factory for other flow rates)
    OXYGEN ANALYSIS METHOD Paramagnetic
    O2 RANGES 0 – 1% up to 0-100% O2 Full Scale, Four Definable Ranges
    O2 RESPONSE TIME T90 < 2 Seconds
    600 SERIES FEATURES:
    OUTPUTS AVAILABLE TCP/IP, RS232, Four Scalable Analog 0-10 V / 4-20 mA
    (Allows Offset & Expandable Range DC Analog Outputs)
    DISCRETE CONTROL
    Remote/Local Control, Range Change, Range Sense
    Mode (All TTL Logic)
    DISCRETE ALARMS
    (Local & Remote Adjustable) General Fault/ TTL Logic (Ground True)
    Calibration Failure/ TTL Logic (Ground True)
    Concentration (2 each)/ TTL Logic (Ground True)
    DIGITAL DIAGNOSTICS Pressure
    Temperatures Pressure Control Voltages
    Flow Parameters
    KEYPAD DISPLAYS Factory Settings
    TCP/IP Address
    Passwords (4) Scalable Analog Output Voltages
    Full Scale Range Select
    Auto Cal Times
    SPECIAL FEATURES Auto Ranging
    Auto Calibration (adjustable through internal clock)
    DISPLAY 3” x 5” Back lit LCD
    SAMPLE TEMPERATURE Up to 50°C, Non-condensing
    AMBIENT TEMPERATURE 5 to 45°C
    AMBIENT HUMIDITY Less than 90% RH Non-condensing
    WARM-UP TIME 1 Hour (Typical)
    FITTINGS 1/4 Inch Tube
    POWER REQUIREMENTS 115/230 (±10%) VAC; 50/60 Hz; 300 Watts Maximum
    DIMENSIONS 5¼ H x 19 W x 23 D (Inches)
    WEIGHT 30-45 Pounds (Depending on configuration)

    Specifications are subject to change without notice.

    pdf Download PDF Brochure