The California Analytical Models 601-2-3 NDIR/O2
digital analyzer
product line is designed around a state of the art 16 bit
microprocessor, with 16 digital inputs, 16 digital outputs, 16 analog
inputs and 4 analog outputs. The analyzer can be manually operated from
the keypad or remotely via TCP/IP, RS-232C communications and discrete
inputs. The analyzer display includes screen presentation of all
analyzer alarms. Four levels of password protection are provided. For
precision measurements, the analyzer’s accuracy is increased
by
entering calibration curve fit polynomials. Automatic calibration may
be activated locally or remotely and includes auto cal via preset
times.
METHOD OF OPERATION
The California Analytical Instruments’ NDIR
analyzer is based on the
infrared absorption characteristics of gases. Using a single infrared
beam to measure gas concentrations, this analyzer produces highly
stable and reliable results. A single infrared light beam is modulated
by a chopper system and passed through a sample cell of predetermined
length containing the gas sample to be analyzed. As the beam passes
through the cell, the sample gas absorbs some of its energy. The
attenuated beam (transmittance) emerges from the cell and is introduced
to the front chamber of a two-chamber infrared microflow detector. The
detector is filled with the gas component of interest and consequently
the beam experiences further energy absorption. This absorption process
increases the pressure in both of the chambers. The differential
pressure between the front and rear chambers of the detector causes a
slight gas flow between the two chambers. This flow is detected by a
mass-flow sensor and is converted into an output signal.
METHOD OF OPERATION - OXYGEN
The
California Analytical oxygen analyzer section utilizes the paramagnetic
method to determine the percent level of oxygen contained in the sample
gas. The oxygen level is displayed on the LCD panel in engineering
units.
FEATURES
Measures from 0-50 ppm up to 100% Full Scale
(CO / CO2 / CH4 / C3H8 / SO2)
Measures Oxygen (O2) from 0–1% up to 100%
Multi-Component
Four User Definable Ranges per Component
Fast Response Time
Auto Ranging / Auto Calibration
Comprehensive Diagnostics
Temp. & Barometric Pressure Compensation
Output Options: Analog (User Scalable), (RS232) using AK
Protocol & TCP/IP
Displays & Outputs: Concentrations, Diagnostics,
Alarms & Preventative Maintenance
Remote Monitoring & Control
APPLICATIONS
Stack Gases (CEM)
Combustion Efficiency
Turbine/Generator Feedback Control
Process Chemical Gas Analysis
Personnel Safety
Fuel Cell Analysis
Vehicle Emission
Model 600P Oxygen Only
Model 601 Any 1 NDIR Component
Model 602 Any 2 NDIR Components
Model 603 Any 3 NDIR Components
Model 601P Any NDIR Component & O2
Model 602P Any 2 NDIR Components & O2
SPECIFICATIONS
IR ANALYSIS METHOD Non-Dispersive Infrared (NDIR)
NDIR COMPONENTS CO / CO2 / CH4 / C3H8 / SO2
DETECTOR TYPE Microflow
NDIR RANGES From 0-50 ppm up to 0-100 % (Dependent upon gas &
cell length)
RANGE RATIO 10:1 (Highest Range/10 = Lowest Range)
RESPONSE TIME (IR) T90 < 2 Seconds to 60 Seconds Adjustable
(Depending on configuration)
IR SAMPLE CELL Stainless Steel with Replaceable Gold Cell Liner
RESOLUTION 0.0001 (Displays Five Significant Digits)
REPEATABILITY Better than 1.0% of Full Scale
LINEARITY Better than 0.5% of Full Scale of Factory Calibrated Ranges
NOISE Less than 1% of Full Scale of Factory Calibrated Ranges
ZERO & SPAN DRIFT Less than 1% of Full Scale per 24 Hours
ZERO & SPAN ADJUSTMENT Via front panel, TCP/IP or RS-232
SAMPLE FLOW RATE 0.25 to 2.0 LPM (Consult factory for other flow rates)
OXYGEN ANALYSIS METHOD Paramagnetic
O2 RANGES 0 – 1% up to 0-100% O2 Full Scale, Four Definable
Ranges
O2 RESPONSE TIME T90 < 2 Seconds
600 SERIES FEATURES:
OUTPUTS AVAILABLE TCP/IP, RS232, Four Scalable Analog 0-10 V / 4-20 mA
(Allows Offset & Expandable Range DC Analog Outputs)
DISCRETE CONTROL
Remote/Local Control, Range Change, Range Sense
Mode (All TTL Logic)
DISCRETE ALARMS
(Local & Remote Adjustable) General Fault/ TTL Logic (Ground
True)
Calibration Failure/ TTL Logic (Ground True)
Concentration (2 each)/ TTL Logic (Ground True)
DIGITAL DIAGNOSTICS Pressure
Temperatures Pressure Control Voltages
Flow Parameters
KEYPAD DISPLAYS Factory Settings
TCP/IP Address
Passwords (4) Scalable Analog Output Voltages
Full Scale Range Select
Auto Cal Times
SPECIAL FEATURES Auto Ranging
Auto Calibration (adjustable through internal clock)
DISPLAY 3” x 5” Back lit LCD
SAMPLE TEMPERATURE Up to 50°C, Non-condensing
AMBIENT TEMPERATURE 5 to 45°C
AMBIENT HUMIDITY Less than 90% RH Non-condensing
WARM-UP TIME 1 Hour (Typical)
FITTINGS 1/4 Inch Tube
POWER REQUIREMENTS 115/230 (±10%) VAC; 50/60 Hz; 300 Watts
Maximum
DIMENSIONS 5¼ H x 19 W x 23 D (Inches)
WEIGHT 30-45 Pounds (Depending on configuration)
Specifications are
subject to change without notice.
|