50 Best Managed Company
Calco Environmental Group recognized by Deloitte as one of Canada's 50 Best Managed Company Award
Product Lines » AMP-Cherokee » so2analyzers
so2analyzers

CEM Monitors - Rental Equipment

API Model 100AH Sulphur Dioxide Analyzer

The API Model100 AH SO2 analyzer used the analytical technique of ultraviolet fluorescence. The instrument is microprocessor controlled and boasts sensitivity down to the 10 ppm range. Ranges are user-defined from 10 to 5,000 ppm.

Manufacturer Advanced Pollution Instrumentation (API)
Analysis Method Ultraviolet Fluorescence
Range 10 to 5,000 ppm
Zero Drift < 20 ppb / 24 hrs.
Span Drift < 1 % of reading / 24 hrs.
Sample Flow Rate 290 cc / minute
Lag Time < 20 secs. switching mode
< 6 secs. NO or NOx only mode
Rise / Fall (95%) 10 secs. NOx only mode
Analog Output 0 - 100 mV, 1, 5, 10 VDC, 4 to 20 mA
Power 115 VAC
Dimensions 17" by 7" by 23.5 "
Weight 44 lbs.

Bovar (Western Research) Model 721ATM Sulphur Dioxide Analyzer

The Bovar Model 721ATM SO2 analyzer uses the analytical technique of ultraviolet fluorescence. The instrument is microprocessor controlled and boasts sensitivity down to the 50 ppm range. Ranges are user-defined from 50 to 5,000 ppm.

Manufacturer Bovar (Western Research)
Analysis Method Ultraviolet Fluorescence
Range 50 to 5,000 ppm
Zero Drift +/- 0.5 % of full scale / 24 hrs.
Sample Flow Rate 1 to 2 liters / minute
Response Time 90 % in < 30 secs.
Sensitivity +/- 5 % of full scale
Analog Output 100 mVDC, 1, 10 VDC, 4 to 20 mA
Power 115 VAC
Dimensions 180 mm by 480 mm by 600 mm
Weight 40 lbs.


Thermo Environmental Model 43H Sulphur Dioxide Analyzer

The Thermo Model 43H SO2 analyzer is a low level analyzer which uses the analytical technique of ultraviolet pulsed fluorescence. The pulsed fluorescence increases sensitivity at lower detection ranges. This analyzer is predominantly used as part of a low level hydrogen sulfide / total reduced sulphur system. Ranging is from 1 to 20 ppm.

Manufacturer Thermo Environmental Instruments
Analysis Method Pulsed Fluorescence
Range 0 - 1, 2, 5, 10, 20 ppm
Zero Drift < 0.1 ppm / 24 hrs.
Span Drift < 1 % of full scale / 24 hrs.
Sample Flow Rate 1 litre / minute
Linearity +/- 1 % of full scale up to 1000 ppm
Analog Output 10 VDC
Power 115 VAC
Dimensions 16.75 " by 8.62 " by 23 "
Weight 45 lbs.

API Model 152 Sulphur Dioxide Analyzer

The API Model 152 SO2 analyzer uses the analytical technique of ultraviolet fluorescence. Three selectable ranges can be chosen. Dilution - Extractive Range Instruments are available.

Manufacturer Advanced Pollution Instrumentation
Analysis Method Ultraviolet Fluorescence
Range 0 - 0.5, 1, 10 ppm
Noise < 1 % of reading
Zero Drift < 0.2 % of reading
Span Drift < 2 % of full scale
Sample Flow Rate 250 cc / minute
Response Time < 60 secs.
Analog Output 0 - 5 VDC
Power 115 VAC
Dimensions 17" by 7" by 22 "
Weight 56 lbs.