Product Lines » 701 analyzer
701 analyzer


Model 701 Ammonia Slip Analyzer 

DESCRIPTION

The CAI Model 701 system consists of an Innova Airtech Model 1312 (1314) Photoacoustic Infrared and a CAI Model 101 Heated sample diluter. The Model 1312 (1314) is an Infrared analyzer with a photoacoustic detector. The analyzer has a dynamic range from 0.07 to 20,000 ppm for ammonia before dilution. Samples containing moisture greater than 1% are appropriately diluted in the 101 heated diluter; and with this analyzer/diluter combination, it is possible to monitor ammonia samples with humidity in excess of 40%. The raw stack sample is drawn by the 101 diluter via heat traced sample lines and a heated internal pump. The appropriately diluted sample is then presented to the 1312 (1314) analyzer for direct analysis.
A critical requirement in the ammonia determination is the transportation of the sample to the analyzer without any losses. Generally, some kind of sample conditioning equipment is used. However, all these processes can result in indeterminate sample losses. As in any extractive technique it is best to transport the sample with heat traced sample lines. Dilution at the stack does not prevent sample losses. The minimum detection limit prior to dilution is 0.07 ppm. The actual minimum detection limit within a sample after dilution depends upon the dilution ratio but is typically down to 0.3 ppm with a 1:5 dilution. This configuration allows the direct analysis of ammonia in samples containing high levels of CO2 and moisture, commonly present in any combustion process. The interference of CO2 in the traditional NDIR ammonia measurements is eliminated by dilution and the cross compensation feature available within the 1312 (1314).

FEATURES

  • Measure five gases simultaneously plus water vapor
  • Easy to use
  • Virtually no maintenance (no carrier gas required)
  • Wide dynamic range
  • Extremely stable, requires only annual calibrations
  • Interference compensation

SPECIFICATIONS

ANALYSIS METHOD NDIR (Photoacoustic)
RANGE 0.07 to 20,000 ppm
RESOLUTION 0.07ppm*
REPEATABILITY 1% of measured value
LINEARITY 0 to 20,00ppm
CO2 EFFECT Negligible
H2O EFFECT Negligible
RESPONSE TIME Step Change 2 to 3 readings
SAMPLE FLOW RATE 2.0 ±0.5 liters/minute
NOISE Less than detection limit
ZERO ±0.2 ppm/month
AMBIENT TEMPERATURE 5 to 45?C
RELATIVE HUMIDITY Less than 90% RH
SAMPLE TEMPERATURE Up to 190?C
SPAN DRIFT ±2.5% of measured value per month
TEMPERATURE DRIFT ±0.3% of measured value / ?C.
SAMPLE PRESSURE 10 to 30 PSIG with the diluter
DILUTION GAS PRESSURE 10 to 30 PSIG
WARMUP TIME 1 hour
FITTINGS 0.25" tube
POWER REQUIREMENTS 115/230 (±) VAC, @ 50/60 Hz; @ 120 Watts
DISPLAY Vacuum fluorescent
DIAGNOSTICS Internal, self diagnostics
OUTPUT RS 232, IEEE 488
Analog optional (4 to 20 mA DC)
Data Storage Capacity 12 days, monitoring 6 gases every 10 minutes
SOFTWARE Windows 95, 98 and 2000
DIMENSIONS 6.9"H x 15.6" W x 11.8" (175mm x 395mm x 300mm)
WEIGHT 19.8 lbs (9 kg)
MAXIMUM PUMPING RATE 30 cc/sec (flushing sampling tube) and 5 cc/sec (flushing measurement chamber)
* Before dilution; the actual value depends upon the degree of dilution

    101 HEATED SAMPLE DILUTER SPECIFICATIONS

    SPAN & DILUENT PRESSURE
    (Electronic Control) 20 TO 30 PSIG
    DILUTION RANGE 1:20
    OVEN TEMPERATURE Factory Set 65 to 190C
    REPEATABILITY 0.1%
    LINEARITY 0.1%
    SAMPLE FLOW RATE 2 to 5 l/m
    PARTICULATE FILTER 0.1 Micron field replaceable
    DISPLAY 3½ Digit panel meter
    DIAGNOSTICS 6 position switch; sample pressure, diluent pressure, ±15V, pump temperature, oven temperature
    ANALOG OUTPUT 0 to 10 VDC; 4 to 20 mA DC; 0 to 20 mA DC
    AMBIENT TEMPERATURE 5 to 45C
    WARMUP TIME 1 Hour
    RELATIVE HUMIDITY Less than 90 % RH; noncondensing
    POWER REQUIREMENTS 115/230 (±10 %) VAC; 60/50 Hz, 750 Watts
    DIMENSIONS 5¼” H x 19” W x 23” D (133mm x 483mm x 508mm)
    WEIGHT 50 lbs