50 Best Managed Company
Calco Environmental Group recognized by Deloitte as one of Canada's 50 Best Managed Company Award
Product Lines » Teledyne » model 100EU
model 100EU

Model 100EU


Download Brochure
UV Fluorescence Low Level SO2 Analyzer

The Model 100EU Low Level SO2 Analyzer uses the proven UV fluorescence principle, coupled with state of the art microprocessor technology to provide accurate and dependable measurements of low level SO2. Exceptional stability is achieved with the use of an optical shutter to compensate for PMT drift and a reference detector to correct for changes in UV lamp intensity.

A hydrocarbon "kicker" and advanced optical design combine to prevent inaccuracies due to interferents. The Model 100EU combines rugged construction, ease of use, powerful diagnostics and outstanding performance to yield the ideal tool for today's air monitoring requirements.

Multi-tasking software allows field range changes as well as auto ranging and dual range operation. The intuitive menu-driven program makes operator interface easy and provides complete accessibility. The Model 100EU can be controlled through the front panel or remotely via RS-232. Built-in diagnostics, electrical and optical tests and warning alarms speed troubleshooting and pinpoint problems.

Features:
  • Versatile 0-50 to 0-20,000 ppb ranges
  • Ranges front panel selectable
  • Microprocessor controlled for versatility
  • Dual ranges and auto ranging
  • Adaptive signal filtering optimizes response time
  • Bi-directional RS-232 ports for remote operation
  • Displays test variables during operation
  • Temperature & pressure compensation
  • Critical orifices provide flow stability
  • Continuous self checking with warning alarms


Specifications:
Ranges: 0-50 ppb to 0-20,000 ppb, user selectable. Dual ranges and auto ranging supported
Units: ppb, ppm, mg/m3
Zero noise: 25 ppt (RMS)
Zero drift: < 0.2 ppb/24 hours
Span drift: < 0.5% of FS/24 hours
Lower Detectable Limit (LDL): 50 ppt
Response time: < 120 seconds to 95%
Sample flow rate: 650 cc/min ±10%
Linearity: 1% of full scale
Operating Temperature Range: 5 - 40°C
Dimensions (HxWxD): 7" (178 mm) x 17" (432 mm) x 23.5" (597 mm)
Weight: 45 lbs (20.5 kg)
Power: 100V 50/60 Hz, 115V 60Hz, 220V 50/60Hz, 230V 50Hz, 240V 50Hz, 250W
Analog Outputs: 10V, 5V, 1V, 100mV, selectable
Current Output: 0-20mA or 0-40mA isolated output, optional
RS-232: Standard, DB-9 connector
Status (Digital): 12 outputs from optoisolator, included with standard configuration
Approvals: CE