Model 700E
Mass Flow Calibrator
The Model 700E is a microprocessor based calibrator for precision gas
analyzers. Using a combination of highly accurate mass flow controllers
and compressed sources of standard gases, calibration standards are
provided for multipoint span and zero checks. Up to 4 gas sources may
be used.
In addition, the Model 700E can be equipped with an optional built-in,
programmable ozone generator for accurate, dependable ozone
calibrations and to produce NO2 when blended
with NO gas in the
internal GPT chamber.
A multi-point linearization curve is used to control the generator to
assure repeatable ozone concentrations, and the optional photometer
allows for even more precise control of the ozone generator, for both
calibrations and Gas Phase Titrations (GPT). To ensure even more
accurate NO2 output, the calibrator with photometer option measures the
ozone concentration prior to doing a GPT.
As many as 50 independent calibration sequences may be programmed into
the M700E, covering time periods of up to one year. The setup of
sequences is simple and intuitive. These sequences may be actuated
manually, automatically, or by a remote signal. The sequences may be
uploaded remotely, including remote editing! All programs are
maintained in non-volatile memory.
The M700E design emphasizes fast response, repeatability, overall
accuracy and ease of operation. It may be combined with the Model 701
Zero Air Generator
to provide the ultimate in easy to use, precise calibration for your
gas analyzers.
Features:
- Advanced E Series electronics
- Lightweight for transportability
- Optional Ethernet connectivity
- Generates precise calibration gases for Ozone, NO, NO2,
CO, HC, H2S, SO2
- 12 independent timers for sequences
- Nested sequences (up to 5 levels)
- Software linearization of Mass Flow controllers
- 4 calibration gas ports configurable for single or
multi-blend gases
- Glass GPT chamber (Gas Phase Titration)
- Ozone generator and photometer allows use as primary or
transfer standard
| Specifications: |
| Dilution
System |
| Flow Measurement
Accuracy: |
± 1%
full scale |
| Repeatability of
Flow Control: |
±
0.2% full scale |
| Linearity of
Flow Measurements: |
±
0.5% full scale |
| Flow Range of
Dilution Air: |
0 to 10 SLPM
Optional Ranges: 0 to 20 SLPM, 0 to 5 SLPM |
| Flow Range of
Cylinder Gases: |
0 to 100 cc/min.
Optional ranges: 0 to 50 cc/min., 0 to 200 cc/min. |
| Zero Air
requirements: |
10 SLPM @ 30 psi
Optional: 20 SLPM @ 30 psi |
| Calibration Gas
Input Ports: |
4 (configurable) |
| Diluent Gas
Input Ports: |
1 |
| Response time: |
60 seconds (98%) |
| Ozone
Generator Module |
| Maximum Output: |
6 ppm LPM |
| Minimum Output: |
100 ppb LPM |
| Response Time: |
180 sec. (98%) |
| Optical Feedback: |
Standard |
| Stability (7
days): |
1% with
photometer option; 3% without photometer option |
| Linearity: |
1% with
photometer option; 3% without photometer option |
| UV
Photometer Option |
| Full Scale Range: |
100 ppb to 10
ppm, user selectable |
| Precision: |
1.0 ppb |
| Linearity: |
1.0% of reading |
| Rise/Fall Time: |
< 20
seconds (Photometer response) |
| Response Time
(95%): |
180 seconds
(System response) |
| Zero Drift: |
< 1.0
ppb/7 days |
| Span Drift: |
< 1.0% /
day, < 2% / 7 days |
| Electrical
Specifications |
| Power
Requirements: |
85 - 264 V, 47 -
63 Hz |
| Test Channel
Analog: |
± 100
mV 1, 5, 10 volts |
| Digital control
outputs: |
12 opto-isolated |
| Digital control
inputs: |
12 opto-isolated |
| Digital status
outputs: |
12 opto-isolated |
| Physical
Specifications |
| Temperature
Range: |
5-40° C |
| Input Pressure: |
20-40 psig
required |
| Weight: |
29 lbs (13.1
kg); 39 lbs. (17.7 kg) including photometer, GPT, and 03 generator |
| Dimensions
(HxWxD): |
7”
(178 mm) x 17” (432 mm) x 24” (609 mm) |
| Operating
Altitude: |
10,000 ft Maximum |
|
|