The PM-10 High Volume Air Sampler is complete with
size selective inlet, outdoor anodized aluminum shelter, aluminum
blower motor assembly, 8" x 10" stainless steel filter holder, mass
flow controller with 20/60 scfm flow probe,7-day mechanical timer,
elapsed time indicator, continuous flow/pressure recorder, and filter
paper cartridge (110v/60hz or 220v/50hz).
|
|
Specifications
|
| Inlet collection efficiency:
Meets EPA's PM-10 Code of Federal Regulations Appendix J to
Part 50 |
| Flow rate: 40
cfm (adjustable) (1.02 to 1.24m³) |
| Filter media:
Micro-Quartz filter media 8" x 10" for PM10 (TE-QMA) |
| Flow control:
MFC (TE-300-310) |
| Motor blower:
2-stage vacuum 0.6 HP 110v/60hz (TE-116311), 220v/50hz (TE-116312) |
| Flow Indicator:
Flow/pressure recorder (TE-5009) with TE-106 Charts |
| Mechanical programmable timer:
7-day mechanical timer (TE-5007) |
| Elapsed time indicator device:
Elapsed Time Indicator (TE-5012) |
Sample Module:
Particulate/vapor sampling module with Tischaloy coating (TE-1002)
|
Shipping Information
|
| Size: 22.5" x
20" x 45.5" sampler carton, 32" x 32" x 25.5" inlet carton |
| Weight: 65
lbs. – sampler carton, 60 lbs. - inlet carton |
| Assembled Space Requirement:
30" round x 64" high |
| Power Requirement:
Single phase with earth ground, 110v/60hz or 220v/50hz |
Motor Specifications:
- 110v/60hz – Part #TE-116311, 2-stage
vacuum 0.6 HP, double ball bearing thru-flow discharge
- 220v/50hz – Part #TE-116312, 2-stage
vacuum 0.6 HP,ball/sleeve bearing thru-flow discharge
|
| Altitude Application:
to 10,000 feet above sea level |
| Ambient Temperature Application:
-30 to + 50ºC |
| Relative Humidity Application:
0 - 100% |
| Sample Delay Start Range:
0 - 7 days |
| Start up amps:
12 amps |
| Running amps:
4 amps |